6.1 Input Signal
SIGNAL
SELECTING
SIGNAL TYPE RANGE RESOLUTION ACCURACY
INPUT
IMPEDANCE
FACTORY
SETTING
B
T/C b 600 ~ 1700°C 1°C 0.5% >100K
Yes
5
T/C S 0 ~ 1700°C 1°C 0.5% >100K
R
T/C R 500 ~ 1600°C 1°C 0.5% >100K
T
T/C T 0.0 ~ 400.0°C 0.1°C 0.5% >100K
E
T/C E 0 ~ 1000°C 1°C 0.5% >100K
J
T/C J 0 ~ 1000°C 1°C 0.5% >100K
T/C K 0 ~ 1200°C 1°C 0.5% >100K
Pt
T/C -200 ~ 600°C 0.1°C 0.5%
0.2mA
Cu50/Cu100 Cu50/Cu100 -50.0 ~ 150°C 0.1°C 0.5% Back
Ln
or
0 ~ 400 Ω
-1999 ~ 9999
12bit A/D
0.5% 0.2mA
0-50mV -20 ~ 100mV 0.5% >100K
0-1V 0.5% >20K
0-5V 0.5% >100K
0-10V
0-10V 0.5% >200K
0-10mA
0.5% <102Ω 4-20mA 0-20mA
4-20mA
Note:
1. The resolution of R/S/B is 1°C; Resolution of the other signals is 0.1 in case the display is value less than 400. In case the
display value more than 400, the resolution is 1°C, over range display or negative display can also be specified at the time of
order..
2. Pulse: Square wave, triangle wave, sine wave (need to mention when order), rating more than 4V or less than IV. Frequency
Range: 0-60KHz.
6.2 Function Keys and Lamps
6.2.1 KEYS
A: SET: Set Key
B: ▲/▼: Increase/Decrease Key
C: SET+▲: Hold the set key, press ▲ will shift digit.
D: LED flashes, press SET key to confirm Parameter.
E: The digit shifts to another one to be modified.
6.2.2 LAMPS
SV red lamp On: Indicate the set value; Flash: Indicate the external setting input value (the second input value of loop 2).
MV red lamp On: Indicate control output value. In case it is manual manipulation, it indicate the manipulate.
MAN green lamp On: Manual manipulate; Flash: In the case of auto-tune; Off: automatism.