Abus-technologies myPCLab Data Acquisition / Recorder Manual de usuario Pagina 8

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 12
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 7
ABUS TECHNOLOGIES INC.
8
myPClab
3.7 Voltage Digital Input Wiring
Digital signals are connected to channel 3, terminals 1 and 2. Observe
indicated polarity. Voltage levels applied to this channel must be within the
specified limits. See “Technical Parameters”.
5. INSTALLATION
5.1 Recommendation
myPCLab inputs are not electrically isolated from the USB port. Before connecting
signals from sensors or transducers, make sure that they are not in contact with any source of
electrical power (wires or conductive surfaces which may get energized). Input channels are not
isolated from each other.
5.2 Input Signal
INPUT TYPE MAXIMUM RANGE
J Thermocouple
-130,0 to 940,0 °C (-202,0 to 1724,0 ºF)
K Thermocouple
-200,0 to 1370,0 °C (-328,0 to 2498,0 ºF)
T Thermocouple
-200,0 to 400,0 °C (-328,0 to 752,0 ºF)
E Thermocouple
-100,0 to 720,0 °C (-148,0 to 1328,0 ºF)
N Thermocouple
-200,0 to 1300,0 °C (-328,0 to 2372,0 ºF)
R Thermocouple
0,0 to 1760,0 °C (-32,0 to 3200,0 ºF)
S Thermocouple
0,0 to 1760,0 °C (-32,0 to 3200,0 ºF)
B Thermocouple
500,0 to 1800,0 °C (932,0 to 3272,0 ºF)
Pt100 RTD
-200,0 to 650,0 °C (-328,0 to 1202,0 ºF)
0 to 50 mVdc Adjustable from -32767 to +32767
0 to 10 Vdc (Channel 1) Adjustable from -32767 to +32767
4 to 20 mAdc Adjustable from -32767 to +32767
myPCLab module sensors and signals
Vista de pagina 7
1 2 3 4 5 6 7 8 9 10 11 12

Comentarios a estos manuales

Sin comentarios